Description
BCS-975R Premium Battery Cyclers
High power, ultra-precise battery cyclers with fully native EIS capability
The BCS-975R is a high-power, regenerative battery cycler. With 8 channels, each capable of 75 A and up to 300 A in parallel, it is purpose built for pouch and prismatic cells.
BioLogic’s Premium Battery Cyclers: High-end cyclers for demanding needs.
Our Premium Battery Cyclers are composed of the BCS-900 Series. This system merges unmatched performance with a user-centric design, owing to the groundbreaking BT-Lab® Suite, your trusted partner in battery testing and data analysis software. This series of ultra-precise cyclers features five current ranges, fast bandwidths (rise and fall times), smart sampling, interfacing with external equipment and temperature probes, amongst many other high-end functions.
The result is a powerful, modular system suitable for any organization specializing in battery testing and research, from industry to academia, quality control to certification.
The BCS-975R Battery Cycler: High-Power Testing for Advanced Research Applications.
The BCS-975R is a regenerative battery cycler designed for pouch and prismatic cell testing and cycling. With native EIS and extendable current up to 300 A, this instrument is designed for high-power performance battery testing. Built with uncompromising performance in mind, the BCS-975R delivers exceptional precision and versatility through its modular design, including:
- 8 channels maximizing your testing capacity
- Up to 75 A per channel
- 3 current ranges for precise control
- Optional Auxiliary temperature monitoring
- BT-Lab® Suite Software (including BT-Test™ & BT-Analysis™)
Specifications
| BCS-975R battery cycler | ||
| Channels per module | 8 | |
| Voltage | ||
| Range | 0 V to 5 V | |
| Control resolution | 100 µV | |
| Measurement resolution | 0.360 µV (24 bit) | |
| Accuracy | 15 mV + 0.05% of setting | |
| Slew rate | 2.5 kV/s | |
| Current | ||
| Max (continuous) per channel | ±75 A; up to ±300 A, with 4 channels in parallel |
|
| Ranges | 3: 75 A down to 750 mA | |
| Control resolution | Down to 360 nA | |
| Measurement resolution | Down to 8.9 nA (24 bit) | |
| Accuracy | 0.03% of FSR* + 0.03% of setting | |
| EIS | ||
| Built-in | Optional on each module | |
| Range | 2 kHz – 10 mHz | |
| Measurement | ||
| Acquisition time | 1 ms | |
| Time base | 5 ms | |
| Additional measurement | ||
| Thermocouple | Optional Auxiliary | |
| Cell connection | ||
| 4 terminal leads | ||
| General | ||
| Height | 16U | |
| Weight | 120 kg | |
| Power consumption | 5000 W | |
*FSR: Full Scale Range