BCS-1012 Battery Cyclers series

BCS-1012 Battery Cyclers series

Up to 6 A – Perfect for cylindrical cells 

Description

BCS-1012  Battery Cyclers series

Consistent performance for high-volume cell aging & validation

BCS-1012 series is a modular battery cycling system designed for high-volume aging studies, offering 32 independent channels with a maximum current of ±6 A per channel.

BCS-1000 Series Battery Cycler: High-Volume Performance for Industrial Environments.

The BCS-1012 is the first model in our Essential BCS Instrument portfolio, purpose-built for high-volume battery testing. Ideal for battery screening and aging studies, the BCS-1012 delivers cost-effective performance without compromising the quality standards you expect from BioLogic.

  • High channel density – Maximizing your testing capacity
  • Up to 6 A – Perfect for cylindrical cells
  • 3 current ranges – For precise control
  • Auxiliary temperature monitoring
  • BT-Lab Suite® Software – includes BT-Test™ & BT-Analysis™

Master aging and validation studies

With industry-leading channel density—up to 128 channels per cabinet, each module providing 32 channels—the BCS-1012 is a robust solution for cylindrical cell testing.

Three current ranges ensure accuracy:

  • 6 A
  • 600 mA
  • 60 mA

The instrument offers a fast data acquisition with a 5 ms sampling rate and time base. Learn more about data of interest to work more efficiently.
Equipped with BCS-CORE, your data is securely and redundantly stored, ensuring reliability in long-term testing. Its modular architecture supports high-volume testing and scalable expansion to match your evolving research needs.
By focusing on core capabilities, the BCS-1012 optimizes cost without compromising the quality and reliability expected from BioLogic instrumentation.

Specifications

BCS-1012
Channels per module 32
Voltage
Range 0 V to 5 V
Control resolution 200 µV
Measurement resolution 0.83 µV (24 bit)
Accuracy 0.5 mV + 0.04% of setting
Slew rate 5 kV/s
Current
Max (continuous) per channel ±6 A
Ranges 3: ±6 A, ±600 mA, ±60 mA
Control resolution Down to 1.2 µA
Measurement resolution Down to 8.9 nA (24 bit)
Accuracy 0.06% of FSR
0.06% of FSR (6 A) + 0.3% of setting
EIS
Built-in No
Measurement
Acquisition time 5 ms
Time base 5 ms
Additional measurement
Thermocouple Optional Auxiliary
Cell connection
4 terminal leads + Ground
General
Height 8U
Weight 36.5 kg
Power consumption 1750 W

*FSR: Full Scale Range – ex. For BCS-1012 at 6 A I-Range the FSR is 12 (-6 A to +6 A).